Experimental assessment of proximal stent-graft (InterVascular) fixation in human cadaveric infrarenal aortas.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 10071619.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper investigates the radial deformation load of an aortic endoluminal prosthesis and determines the longitudinal load required to cause migration in a human cadaveric aorta of the endoprosthesis. The endovascular prosthesis under investigation was a 24 mm diameter, nitinol, self-expanding aortoaortic device (InterVascular, Clearwater, Florida, U.S.A.). Initially, a motorised digital force gauge developed an incremental load which was applied to the ends of five stent-grafts, to a maximum of 10 mm (42%) compression. Secondly, using a simple bench model, each ends of four stent-grafts were deployed into 10 cadaveric experimental aneurysm necks and a longitudinal load applied to effect distraction. Increasing load produced increasing percentage deformation of the stent-grafts. The mean longitudinal distraction load for an aneurysm neck of 20 mm was 409 g (200-480 g), for 15 mm was 277 g (130-410 g) and for 10 mm was 218 g (130-340 g). The aneurysm diameter and aortic calcification had p values of 0.002 and 0.047, respectively, while the p value for aneurysm neck length was less than 0.00001. These results suggest that there is a theoretical advantage of oversizing an aortic prosthesis and that sufficient anchorage is achieved in an aortic neck of 10 mm to prevent migration when fully deployed.
Medical subject headings
- Aorta, Abdominal
- Blood Vessel Prosthesis Implantation
- Stents