A paired wedge filter system for compensation in dose differences.

Kobayashi, H; Sakurai, Y; Kondo, S; Abe, S; Hayakawa, N; Aoyama, Y; Obata, Y; Ishigaki, T · Radiother Oncol · 1998

other · Level V

Where this comes from

Abstract

In radiotherapy, it is important to conform the high dose volume to the planned target volume. A variable thickness paired wedge filter system was developed to compensate for dose inhomogeneity arising from field width segment variation in conformal irradiation. The present study used a 6 MV linear accelerator equipped with multileaf collimator leaves and a paired wedge compensating filter system. The dose variation due to field width was measured in each field segment width. The variation in attenuation of the compensators was measured as a function of filter position. As the field width increases, the relative absorbed dose also increases; this is the point of requiring compensation, so it can be in reverse proportion. As the field width increases, the relative absorbed dose also increases; this is why compensation is required and thus it must be in reverse proportion. Attenuation of the absorbed dose by the paired filters was in proportion to the filter position. The filter position to compensate for the difference of absorbed doses was defined by the square root of the field width. For a field varying in width from 4 to 16 cm, the variation in the absorbed dose across the field was reduced from 12% to 2.7%. This paired wedge filter system reduced absorbed dose variations across multileaf collimator shaped fields and can facilitate treatment planning in conformal therapy.