Evaluation of the cushion setting on performance of a Biodex II dynamometer.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 10084449.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the effects of the soft cushion setting with those of the hard cushion setting of a Biodex II isokinetic dynamometer on peak velocity and peak torque with six loads at 23 preset speeds. Mixed within- and between-groups, repeated-measures design. Research laboratory of a physical therapy education program. Each load was strapped onto the right knee attachment of the dynamometer and was dropped through a 100 degrees arc of motion five times at each preset speed using the soft and hard cushion settings. Peak velocity, peak torque, time to peak velocity, time to peak torque. Peak velocity was significantly influenced by the cushion setting, the weight dropped, and the preset speed. Maximal peak velocity achieved was 400 degrees/sec. Peak torque was influenced significantly by the cushion setting, the load, and the preset speed. Maximal peak torque measured was 34.99 kg-m. The rotating shaft of Biodex II dynamometer moved at a slower speed through the entire arc of motion at the soft cushion setting and often failed to achieve a constant, preset speed. Peak torque measured using the soft cushion setting was more variable and therefore questionable.