Volume flow estimation by colour duplex.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 10092894.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the accuracy of volume blood flow using a digitised colour duplex scanner. Observer-blinded experimental study. Method comparison was performed with linear regression analysis of 89 paired observations in 11 anaesthetised pigs. A Siemens Sonoline Elegra ultrasound system was used for transcutaneous volume flow estimation using invasive transit time flowmetry by Cardiomed as a reference. For the individual measurement we found a standard error of the estimate (SEE) of 22 ml/min. For the regression line, however, the SEE was only 0.2 ml/min. Digitised colour-duplex sonography has a volume flow measurement error that is too high for single measurements in the individual patient for the method to be useful in clinical decision making, but sufficient for examinations of groups and comparison of groups.
Medical subject headings
- Blood Flow Velocity
- Blood Volume
- Ultrasonography, Doppler, Color