Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 10220194.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty (TKA) has not been well reported before. A 63-year-old man with osteoarthritis of the right knee underwent TKA with a New Jersey LCS Knee, with cementing on the tibia and patella but not on the femoral component. After 42 months, in addition to wearing of polyethylene of the tibia and patella, severe osteonecrosis of the medial femoral condyle was noted. Osteonecrosis caused loss of osseous support of the medial flange of the femoral component, and the bone ingrowth of the central and lateral flange to the distal femur was so good that it overcame the yield stress of the metal of the femoral component and caused fracture of the femoral component. The osteolytic area was filled with autogenous iliac bone, and a new femoral component was inserted and cemented. The patient's condition became satisfactory with relief of pain. Although uncommon, fracture of the femoral component does occur associated with polyethylene wear and osteolysis.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Biocompatible Materials
- Knee Prosthesis
- Osteolysis
- Polyethylenes
- Prosthesis Failure
Anatomy
- femur
- knee