Biocompatibility of a silicone-coated polypropylene hollow fiber oxygenator in an in vitro model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10355404.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A silicone-coated microporous hollow-fiber membrane oxygenator has been developed to prevent plasma leakage during long-term use. The objective of this study was to evaluate the biocompatibility of the oxygenator. A silicone-coated oxygenator was compared with an uncoated oxygenator in an in vitro model of cardiopulmonary bypass. Simulated circulation was maintained for 6 h at 37 degrees C. Platelet counts decreased significantly (p < 0.05) and leukocyte counts tended to decline; however, the differences between groups were not significant. Concentrations of C3a increased significantly in both groups (p < 0.05), but levels were significantly less in the silicone-coated oxygenator (p = 0.008). In contrast, concentrations of C4a, beta-thromboglobulin, and granulocyte elastase increased significantly (p < 0.05), but the differences between groups were not significant. Silicone coating over a microporous hollow-fiber membrane may improve biocompatibility by reducing C3a activation.
Medical subject headings
- Coated Materials, Biocompatible
- Complement C3a
- Extracorporeal Circulation
- Oxygenators, Membrane
- Silicones