Interface pressures and shear stresses: sagittal plane angular alignment effects in three trans-tibial amputee case studies.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10355640.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Interface pressures and shear stresses at different sagittal plane angular alignment settings were measured on 3 trans-tibial amputee subjects ambulating with patellar-tendon-bearing total contact prostheses. Substantial socket-shank angular alignment modifications in the sagittal plane had minimal effect on stance phase peak interface pressures, though more substantial effects on stance phase peak resultant shear stresses. No consistent trend of a greater stress at misaligned vs nominally aligned settings was identified. Changes in interface stresses from session to session tended to be greater than those for different alignment settings, suggesting that subjects compensated well for misalignments but less well for session differences.
Medical subject headings
- Amputation Stumps
- Artificial Limbs
- Prosthesis Fitting
- Stress, Mechanical
Anatomy
- tibia