Percutaneous trigger finger release: a comparison of a new push knife and a 19-gauge needle in a cadaveric model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10447182.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An instrument was developed for percutaneous release of the A1 pulley and this procedure was compared with percutaneous release using a 19-gauge needle in 20 fresh-frozen cadaver hands. This instrument was developed to avoid the complications of incomplete release and flexor tendon damage previously reported with percutaneous release using a 19-gauge needle. Seventy-eight fingers were used in the study. The push knife was used to release 52 fingers. The A1 pulley was completely released in 51 fingers (98%) and there were no incidents of flexor tendon damage. A 19-gauge needle was used to release 26 fingers. The A1 pulley was completely released in only 10 fingers (38%) and the flexor tendon was damaged in 19 (73%). In cadaveric hands, this instrument proved to be more effective at creating a complete release of the A1 pulley and demonstrated less disruption of the flexor tendon surface compared with current needle techniques.
Medical subject headings
- Fingers
- Needles
- Surgical Instruments
- Tendons
- Tenosynovitis