Radiation safety parameters following prostate brachytherapy.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10487562.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the degree and variability of radiation exposure to the general public from patients after I-125 or Pd-103 prostate brachytherapy. Radiation exposure measurements were made from 38 consecutive, unselected patients with stage T1 or T2 prostatic carcinoma who had transperineal I-125 or Pd-103 implants at the University of Washington in 1998. The exposure rate at the anterior skin surface following a I-125 implant ranged from 2.2 to 8.9 mrem/hour (average: 5.0). The exposure rate at the anterior skin surface from a Pd-103 implant ranged from 0.5 to 4.9 mrem/hour (average: 1.7). Based on the current Nuclear Regulatory Commission (NRC) regulations the time required to reach the annual limit at the anterior skin surface would be 20 hours for I-125 and 59 hours for Pd-103. For exposure at the lateral skin surface, the times would exceed 500 hours for either isotope. This data suggest that patients need not be concerned about being a radiation risk to the general public following their procedure.
Medical subject headings
- Brachytherapy
- Carcinoma
- Environmental Exposure
- Prostatic Neoplasms