In vivo hemodynamic evaluation based on transverse Doppler measurements of blood velocities and vessel diameter.

Wan, M; Gong, X; Qian, M · IEEE Trans Biomed Eng · 1999

basic_science · Level V

Where this comes from

Abstract

We have developed a method for simultaneously measuring the blood flow velocity distributions along a line perpendicular to the vessel wall and the instantaneous diameter of the common carotid artery. Thus, the wall shear rates and shear rate distribution across the arterial lumen could be computed accurately. In this method, transverse Doppler technique is combined with ultrasonic A echo-mode, using only one single crystal transducer. In order to satisfy the requirements of high spatial and velocity resolutions, especially extending to a very low velocity range in the case of low signal-to-noise ratio, a combination scheme of 10-MHz operating frequency, sample volume tracking, 12-bit A/D, 512-point fast Fourier transform with 5-kHz sample rate, spectrum average, variable threshold for bandwidth estimation, smoothing of velocity curves, and hardware was carefully designed. The velocity distribution measurements based on transverse Doppler spectrums were also validated under the steady flow conditions using a flow loop system. Results from this study show this ultrasonic system to be capable of measuring the shear rates at wall accurately and shear rate distributions across the common carotid arterial lumen.