Comparison of esophageal Doppler monitor generated minute distance and cardiac output in a porcine model of ventricular fibrillation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10507713.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The primary goal of cardiopulmonary resuscitation (CPR) is to increase cardiac output (CO), providing adequate tissue perfusion and oxygenation to maintain normal organ function. A non-invasive, easy to use, commercially available esophageal doppler monitor (EDM, Deltex) has been found to provide minute distance (MD), which is the distance moved by a column of blood through the aorta in 1 min. The goal of our study was to determine if CO measurements correlate with the EDM MD, before and during cardiac arrest, in a porcine model of ventricular fibrillation. Twenty pigs were anesthetized and an EDM was placed. MD measurement using EDM, and CO measurement using florescent microsphere injections were compared before and during CPR. MD correlated well with CO (r2 = 0.96) before and during CPR. Based on the excellent correlation between MD as determined by EDM and CO by florescent microsphere technique, it appears that the non-invasive use of the EDM may play a valuable role in determination of CO during CPR.
Medical subject headings
- Cardiac Output
- Cardiopulmonary Resuscitation
- Heart Arrest
- Ultrasonography, Doppler
- Ventricular Fibrillation