Accuracy of centre of pressure measurement using a piezoelectric force platform.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 10521614.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the accuracy in determining centre of pressure when using a force platform for stabilometry research. Technical report. Previous studies have determined the accuracy of locating the centre of pressure by applying force through a single point. In human stabilometry research however, force is distributed across two feet and the centre of pressure located between the feet. Forces were applied to a Kistler force platform using a force applied through a single point as well as when force was distributed between two metal blocks. Errors in calculating centre of pressure were substantially larger when force was applied through a single point than when the force was distributed between two blocks. Evaluating centre of pressure accuracy by applying force through a single point overestimates the errors inherent in human stabilometry research. Evaluating changes in the centre of pressure of human subjects standing on a force platform is a common tool for evaluating control mechanisms of balance. Accurate information regarding the centre of pressure is essential for the validity of this research.
Medical subject headings
- Postural Balance