Performance characteristics of three serum iron and total iron-binding capacity methods in acute iron overdose.
Level III
Where this comes from
- Record sourced from PubMed, PMID 10549253.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Accurate serum iron and total iron binding capacity (TIBC) measurements may be useful in acute iron overdoses. Two alumina column TIBC methods were found to measure increased TIBC when free iron was present. A homogeneous TIBC method gave consistent results until iron concentrations exceeded 500 micrograms/dL (90 mumol/L), when it began to underestimate the TIBC. Serious iron overdoses require chelation therapy with deferoxamine. Iron recovery was reduced by up to 50% for all 3 methods with clinically achievable concentrations of deferoxamine 8,400 micrograms/dL (150 mumol/L). TIBC measurements by both alumina column methods were reduced by deferoxamine in the presence of free iron and unaffected when the iron concentration was less than the TIBC. The homogeneous TIBC method yielded falsely elevated results in the presence of free deferoxamine. Procedures that measure TIBC by addition of excess ferric iron followed by alumina adsorption are not suitable for monitoring TIBC in acute iron overdose. The homogeneous TIBC assay can be used in acute iron overdose but underestimates TIBC when iron concentrations exceed 500 micrograms/dL (90 mumol/L). None of the methods examined are useful for measuring iron or TIBC in the presence of deferoxamine.
Medical subject headings
- Iron
- Poisons
- Toxicology