Loads on internal spinal fixators measured in different body positions.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10552317.
- Also identified by PMC identifier 3611199.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Telemeterized internal spinal fixation devices were implanted in ten patients. The loads acting on the fixators were compared for different body positions, including standing, sitting, and lying in a supine, prone, and lateral position. Implant loads differed considerably from patient to patient depending, for example, on the indication for surgery and the surgical procedure. They were altered by anterior interbody fusion. Mostly, only small differences in implant loads were found for the various lying positions. Flexion bending moments were significantly higher in upright than in lying body positions. Loads on the fixators were not higher for sitting than for standing. Patients who have undergone mono- or bisegmental spine stabilization should therefore be allowed to sit as soon as they can leave the bed.
Medical subject headings
- Orthopedic Fixation Devices
- Posture
- Telemetry
- Weight-Bearing