A novel method to reduce device-related infections in patients supported with the HeartMate device.

Arusoglu, L; Koerfer, R; Tenderich, G; Alexander, W A; El-Banayosy, A · Ann Thorac Surg · 1999

case_series · Level IV

Where this comes from

Abstract

Improved implantable left ventricular assist device technology has made survival to heart transplantation a near certainty. Nevertheless, infection remains a major risk to recipients of current percutaneous systems. We developed a modified implantation technique applied to the last 9 of 30 patients who received the HeartMate vented electric left ventricular assist system (LVAS). Covering the upper surface of the pump with a patch of knitted graft material was followed by a decline in the incidence of pocket infections from 33.3% to 11.1%. This modification compares favorably to that of a lengthened percutaneous driveline tunnel in reducing device-related infection.