Radiation doses delivered to radiologists during contrast-enhanced helical-CT examinations.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10602977.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This work monitors the radiation doses to a radiologist during supervision of automatic contrast media injections during helical-CT examinations of the chest and abdomen. Forty consecutive standard helical-CT examinations of adult's chest and/or abdomen were monitored with five dosimeters worn by the radiologist supervising the entire injection with the hand on the injection site. Mean doses per examination measured at chest, thyroid gland, and hand levels were 11, 16, and 130 microGy, respectively, during chest examinations, and 5, 7, and 55 microGy during abdominal examinations. According to the high number of CT examinations performed, wearing lead apron, special lead glove protection, and thyroid shield is highly recommended.
Medical subject headings
- Contrast Media
- Occupational Exposure
- Radiation Dosage
- Radiology
- Tomography, X-Ray Computed