A fenestrated covered suprarenal aortic stent.

Browne, T F; Hartley, D; Purchas, S; Rosenberg, M; Van Schie, G; Lawrence-Brown, M · Eur J Vasc Endovasc Surg · 1999

basic_science · Level V

Where this comes from

Abstract

to address the clinical problem of inadequate neck length of abdominal aortic aneurysms in endoluminal surgery. a covered suprarenal aortic stent was designed with a fenestration to preserve blood flow in a targeted renal artery. a Dacron-covered stent was accurately cut to size with a fenestration according to pre-imaging studies. the stent was successfully deployed in canine models, preserving the visceral and renal artery of interest. by accurately placing a covered stent in the aorta and preserving the blood flow to its branches, it may be possible to extend the indications for endoluminal aortic aneurysm grafting.

Medical subject headings