A transconductance driven-right-leg circuit.

Spinelli, E M; Martínez, N H; Mayosky, M A · IEEE Trans Biomed Eng · 1999

other · Level V

Where this comes from

Abstract

Biopotential measurements are very sensitive to electromagnetic interference (EMI). EMI gets into the acquisition system by many ways, both as differential and common mode signals, driven-right-leg circuits (DRL) are widely used to reduce common mode interference. This paper reports an improvement on the classic DRL. The proposed circuit uses a transconductance amplifier to drive the patient's body. This configuration has some interesting properties, which provide an extended bandwidth for high-frequency EMI rejection (such as fluorescent lights interference). The improvement is around 20 dB for frequencies of few kilohertz and the circuit is easy to compensate for stability. A comparative analysis against a typical DRL is presented, the results obtained have been experimentally tested.