Comparison of left and right atrial epicardial pacing in patients with congenital heart disease.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 10617023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Complex congenital heart disease (CHD) often necessitates the use of epicardial pacing. Adequate right atrial (RA) sensing and pacing thresholds are often difficult to obtain due to suture line scarring and RA dilatation. The purpose of this study was to evaluate the placement of left atrial (LA) epicardial leads in children. Patient demographics, pacing, and sensing data of atrial pacing systems implanted between January 1994 and January 1997 were collected. Forty-nine pacing systems were implanted: 14 LA epicardial, 19 RA epicardial, and 16 transvenous in the right atrium. Lead impedance, current, and energy were similar in the two epicardial groups throughout the study. Energy thresholds (ET) were lower in the LA than RA at 6 months, and 1 and 2 years (p < 0.05). Analysis of post-Fontan patients performed alone revealed a lower ET in the LA as compared with the RA. Pacing and sensing parameters from transvenous leads are presented for relative comparison. Transvenous leads are most efficient but often contraindicated in complex CHD. LA leads offer lower energy thresholds than RA leads with similar sensing parameters.
Medical subject headings
- Heart Defects, Congenital
- Pacemaker, Artificial