Abdominal aortic aneurysm measurements for endovascular repair: intra- and interobserver variability of CT measurements.

Aarts, N J; Schurink, G W; Schultze Kool, L J; Bode, P J; van Baalen, J M; Hermans, J; van Bockel, J H · Eur J Vasc Endovasc Surg · 1999

cross_sectional · Level IV

Where this comes from

Abstract

to evaluate the intra- and interobserver variability in measurements of the aorta and iliac arteries in patients with abdominal aortic aneurysms (AAAs) considered for endovascular repair using computed tomography angiography (CTA). the diameter of the neck, aneurysm, right and left iliac artery were measured by 5 observers in 10 consecutive patients. Measurements were performed on hard copy using a ruler and on a workstation using an electronic caliper. the intraobserver variability showed a decrease going from hard copy to workstation in the standard deviation of the differences of the paired observations for the neck from 3.54 mm to 1.18 mm; for the aorta from 4.16 to 1.72 mm; for the right iliac from 1.87 to 1.01 mm; for the left iliac from 2.07 to 0.87 mm. The interobserver variability showed a similar decrease for the neck in all ten pairs of observers; for the aorta in two, for the right iliac and left iliac in five. However, the difference between observers regularly exceeded 2 mm. the use of a workstation and electronic calipers results in lower intra- and interobserver variability. However, the results still show a clinically relevant difference between the observers. Therefore, it is necessary to develop an automatic observer-independent measurement technique.

Medical subject headings