Diamond detector measurements near simulated air channels for narrow photon beams.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10665794.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the combined effect of increased photon transmission, reduced photon scatter, increased secondary electron range and loss of electronic equilibrium for narrow 6-MV beams in and around a simulated air channel. A measuring method was developed in-house for relative dose measurements near simulated air-like/soft-tissue interfaces in an automated water phantom. A Styrofoam cylinder (density 0.03 g/cm3) of 2-cm diameter was submersed in the water phantom and irradiated with small rectangular radiation fields. The field length was fixed at 10 cm and the field widths ranged from 1 to 4 cm. The axis of the foam cylinder and the long side of the field were parallel. A water layer of 2 cm was realised upstream of the cylinder. Relative depth dose and profiles behind the foam cavity were assessed using a diamond detector with a sensitive crystal thickness of 0.21 mm located at 1 mm from the top of the encapsulation. The dose at central axis 1.1 mm behind the cavity was found to be 92 and 74% for a field size of 10 x 2 and 10 x 1 cm2, respectively. The highly convex dose profile of the 10 x 1-cm2 field, characterising the homogeneous case, is flattened. The diamond detector is an excellent choice as a detector in small photon fields with high-dose gradients as they occur near air channels, provided the orientation of the detector is appropriate. Doses near air channels are subject to significant local variations as a function of small changes of field width, and local underdosing may occur in particular cases.
Medical subject headings
- Photons
- Radiotherapy, High-Energy