Xenon-enhanced computed tomography quantifies normal maxillary sinus ventilation.
Where this comes from
- Record sourced from PubMed, PMID 10699821.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the applicability, safety, and normal parameters of a xenon-enhanced CT technique to quantify maxillary sinus ventilation. Nine healthy subjects inhaled a xenon-oxygen-air mixture through their noses while repeated CT scans were performed through the same section of their sinuses. Images were obtained every 1 to 3 minutes and analyzed to measure the density of the gas in the maxillary sinus as a function of time. Individual nasal cavity time constants ranged from 0.5 to 18 minutes. Studies performed after decongestion showed poorer sinus ventilation. The xenon-CT washin/washout technique is safe, effective, and gives representative data.
Medical subject headings
- Contrast Media
- Maxillary Sinus
- Oxygen
- Pulmonary Ventilation
- Radiographic Image Enhancement
- Tomography, X-Ray Computed
- Xenon