Frequency doubling technology perimetry for detection of glaucomatous visual field loss.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 10704546.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the ability of frequency doubling technology perimetry to detect early, moderate, and advanced glaucomatous visual field loss. In a prospective study, frequency doubling technology perimetry (C-20 full threshold) was performed in the right eye of 254 normal control subjects and 230 patients with early (n = 85), moderate (n = 114), or advanced (n = 31) glaucomatous visual field loss. Previous Humphrey Field Analyzer test results were used to classify glaucomatous visual field loss as early (mean deviation no worse than -6 dB), moderate (mean deviation between -6 and -12 dB) or advanced (mean deviation between -12 and -22 dB). Receiver operating characteristic curves showed 100% sensitivity and specificity (area under the curve, 1.0) for detecting advanced glaucomatous visual field loss, approximately 96% sensitivity and 96% specificity (area under the curve, 0.9751) for detecting moderate glaucomatous visual field loss, and approximately 85% sensitivity and 90% specificity (area under the curve, 0.9261) for early glaucomatous visual field loss. Frequency doubling technology perimetry demonstrates high sensitivity and specificity for detection of early, moderate, and advanced glaucomatous visual field loss.