Low-friction arthroplasty with Boneloc bone-cement: outcome at 2 to 4 years.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 10708087.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report the clinical and radiologic outcome of 109 Chamley low-friction arthroplasties implanted with Boneloc bone-cement (Biomet, Bridgend, South Wales, UK) into 104 patients. The mean follow-up was 30 months (range, 2-48 months). There were 72 women (mean age, 71 years) and 32 men (mean age, 72). Cartridge-packed cement was used in 37 cases and vacuum-packed cement in 72 cases. Survivorship analysis based on revision for aseptic loosening showed 79% survival at 4 years. Seventeen (15.5%) hips have been revised for aseptic loosening to date, in which all stems and 4 cups were loose. Extensive femoral osteolysis was always present and resulted in 4 cases of femoral cortical perforation at revision. Survivorship analysis based on revision and radiologic failure showed only 55% survival over the same period. When radiologic loosenings were included as failures, the vacuum-packed cement performed significantly worse than the cartridge-packed cement it replaced. These poor results were consistent with the withdrawal of Boneloc from clinical use in 1995, and we recommend indefinite follow-up for surviving prostheses.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Bone Cements
- Methacrylates
Anatomy
- hip