Coronary physiology revisited : practical insights from the cardiac catheterization laboratory.
Where this comes from
- Record sourced from PubMed, PMID 10725297.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Various coronary physiological measurements can be made in the cardiac catheterization laboratory using sensor-tipped guidewires; they include the measurement of poststenotic absolute coronary flow reserve, the relative coronary flow reserve, and the pressure-derived fractional flow reserve of the myocardium. Ambiguity regarding abnormal microcirculation has been reduced or eliminated with measurements of relative coronary flow reserve and fractional flow reserve. The role of microvascular flow impairment can be separately determined with coronary flow velocity reserve measurements. In addition to lesion assessment before and after intervention, emerging applications of coronary physiology include the determination of physiological responses to new pharmacological agents, such as glycoprotein IIb/IIIa blockers, in patients with acute myocardial infarction. Measurements of coronary physiology in the catheterization laboratory provide objective data that complement angiography for clinical decision-making.