Positron emission mammographic instrument: initial results.

Murthy, K; Aznar, M; Bergman, A M; Thompson, C J; Robar, J L; Lisbona, R; Loutfi, A; Gagnon, J H · Radiology · 2000

basic_science · Level V

Where this comes from

Abstract

Performance characteristics of a positron emission mammographic (PEM) instrument were studied. This dedicated metabolic breast imaging system has spatial resolution of 2.8-mm full width at half maximum (FWHM), coincidence resolving time of 12-nsec FWHM, and absolute efficiency of 3%. Hot spots with diameter of 16 mm in a phantom with signal-to-background activity ratio of 6:1 were distinguishable with a scanning time of 5 minutes.

Medical subject headings