Correlation of esophageal conductance measurements with aortic and left ventricular diameters and stroke volume.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10763303.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Esophageal conductance measurements were correlated with hemodynamic events in 9 dogs chronically instrumented for measurement of left ventricular (LV) and aortic pressures, LV short axis and descending aortic diameters, and aortic blood flow. A four-electrode conductance catheter was positioned in the esophagus. Both an internal and an internal/external configuration were examined during anesthesia with hemodilution, pulmonary lavage and dobutamine infusion. LV stroke volume was altered by caval occlusion at each intervention. Stroke conductance was highly correlated to aortic or LV diameters and stroke volume over a range of diameters depending on the electrode configuration. Esophageal conductance measurements are directly influenced by local hemodynamic events adjacent to the site of measurement.
Medical subject headings
- Aorta, Thoracic
- Esophagus
- Models, Cardiovascular
- Stroke Volume
- Ventricular Function, Left