Femur BMD bias associated with DPX-IQ reanalysis of DPX-acquired scan files.
Where this comes from
- Record sourced from PubMed, PMID 10793876.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bone mineral density values calculated by DPX-IQ software reanalysis of original DPX software analyzed scans were compared at the femur region. Based on the results for 24 subjects each scanned on 12 different occasions, a significant bias at all sites--neck, Ward's triangle and trochanter--ranging from -0.7% (neck) to -2.1% (trochanter) was found. For the trochanter site 16 of the 24 subjects had significantly lower mean BMD values after DPX-IQ reanalysis, a trend also noted at the neck and Ward's triangle. Although it has been recommended that scans analyzed with DPX software do not require DPX-IQ reanalysis, our findings strongly indicate that DPX-IQ reanalysis is essential to minimize total error during follow-up BMD determinations.
Medical subject headings
- Absorptiometry, Photon
- Bone Density
- Femur
- Software