Detection of positional airway obstruction in neonates by acoustic reflection.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10806183.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In neonates intubated with an uncuffed endotracheal tube (ETT), positional changes of the head may induce obstruction (side position-related ETT obstruction [SPRO]) due to abutment of the beveled distal ETT orifice against the tracheal wall. We studied whether the acoustic reflection (ACR) method, a 4-s measurement that maps cross-sectional area as a function of the distance along the ETT and the airways, could detect SPRO. Eleven preterm newborns intubated with 2.5-mm ETTs and clinically suspected of having SPRO were studied with the head oriented to the left and to the right. In all patients there was a marked decrease in the ACR-measured area beyond the distal tip of the ETT in the presence of obstruction (decrease = 38 +/- 22% [mean +/- SD] of the ETT inside area), while the ACR-measured area increased markedly in the absence of obstruction (increase = 49 +/- 17%). For six of the 11 infants, we also recorded the maximal flow produced by a set mechanical inflation pressure. This maximal flow decreased in the presence of obstruction (decrease = 47 +/- 18%), and was constantly associated with a decrease in ACR-measured area beyond the ETT. In conclusion, ACR measurement is an efficient method for diagnosing positional ETT obstruction in intubated newborns.
Medical subject headings
- Airway Obstruction
- Intubation, Intratracheal