A continuous pure moment loading apparatus for biomechanical testing of multi-segment spine specimens.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 10807999.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An apparatus is described that enables the application of continuous pure moment loads to multi-segment spine specimens. This loading apparatus allows continuous cycling of the spine between specified flexion and extension (or right and left lateral bending) maximum load endpoints. Using a six-degree-of-freedom load cell and three-dimensional optoelectronic stereophotogrammetry, characteristic displacement versus load hysteresis curves can be generated and analyzed for different spinal constructs of interest. Unlike quasi-static loading, the use of continuous loading permits the analysis of the spine's behaviour within the neutral zone. This information is of particular clinical significance given that the instability of a spinal segment is related to its flexibility within the neutral zone. Representative curves for the porcine lumbar spine in flexion-extension and lateral bending are presented to illustrate the capabilities of this system.
Medical subject headings
- Lumbar Vertebrae
- Weight-Bearing