Frameless computer-aided surgery system for revision endoscopic sinus surgery.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 10828791.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To increase the intraoperative safety factor and to acquire anatomic assistance during revision endoscopic sinus surgery (RESS), we used an optical computer-aided surgery (CAS) system that we developed collaboratively in Bern, Switzerland. During 1 year, 25 RESSs were performed with CAS: recurrent polyposis (n = 20), recurrent frontal recess stenosis (n = 3), and recurrent frontal recess stenosis with mucocele (n = 2). These patients were compared with a control group of 10 patients undergoing RESS without CAS. The same surgeon (M.C.) performed all operations, and there were no minor or major complications in either group. The clinical inaccuracy of our system is between 0.5 and 2 mm with paired-point and surface matching. The navigation system is an important aid to surgeons in identifying anatomic landmarks that are typically difficult to visualize in this type of surgery, thus reducing the stress placed on the surgeon.
Medical subject headings
- Endoscopy
- Nasal Polyps
- Otorhinolaryngologic Surgical Procedures
- Paranasal Sinus Diseases
- Therapy, Computer-Assisted