Carbon dioxide absorbents containing potassium hydroxide produce much larger concentrations of compound A from sevoflurane in clinical practice.

Yamakage, M; Yamada, S; Chen, X; Iwasaki, S; Tsujiguchi, N; Namiki, A · Anesth Analg · 2000

rct · Level II

Where this comes from

Abstract

We investigated the concentrations of degraded sevoflurane Compound A during low-flow anesthesia with four carbon dioxide (CO(2)) absorbents. The concentrations of Compound A, obtained from the inspiratory limb of the circle system, were measured by using a gas chromatograph. In the groups administered 2 L/min fresh gas flow with 1% sevoflurane, when the conventional CO(2) absorbents, Wakolime(TM) (Wako, Tokyo, Japan) and Drägersorb(TM) (Dräger, Lübeck, Germany), were used, the concentrations of Compound A increased steadily from a baseline to 14.3 ppm (mean) and 13.2 ppm, respectively, at 2 h after exposure to sevoflurane. In contrast, when the other novel types of absorbents containing decreased or no potassium hydroxide/sodium hydroxide, Medisorb(TM) (Datex-Ohmeda, Louisville, CO) and Amsorb(TM) (Armstrong, Coleraine, Northern Ireland), were used, Compound A remained at baseline (<2 ppm) throughout the study. In the groups administered 1 L/min fresh gas flow with 2% sevoflurane, Wakolime(TM) and Drägersorb(TM) produced much larger concentrations of Compound A (35.4 ppm and 34.2 ppm, respectively) at 2 h after exposure to sevoflurane. Medisorb(TM) showed measurable concentrations of Compound A (8.6 ppm at 2 h), but they were significantly smaller than those produced by the two conventional absorbents. In contrast, when Amsorb(TM) was used, Compound A concentrations remained at baseline throughout the study period. Carbon dioxide absorbents containing potassium hydroxide/sodium hydroxide produce much larger concentrations of Compound A from sevoflurane in clinical practice. An absorbent containing neither potassium hydroxide nor sodium hydroxide produces the smallest concentrations of Compound A.

Medical subject headings