Pressure signal transmission of five commercially available oesophageal balloon catheters.
other
Where this comes from
- Record sourced from PubMed, PMID 10872140.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Five commercially available oesophageal balloon catheters (OBCs) were tested to evaluate the accuracy in transmitting fast-changing pressure signals which can be observed, for example, during phrenic nerve stimulation. Research laboratory of a university hospital. The OBCs tested varied in length (900-1390 mm) and inner diameter (0.9-1.5 mm) as well as in balloon material [latex or polyvinylchloride (PVC)]. A 180-cm tube served as a control. A sudden pressure drop was generated by the explosion of a pressurized latex balloon. The time between the pressure drop and 75, 50, 25 and 10% of the maximal pressure was measured. The time intervals required to transduce a pressure drop of 90% varied between the different OBCs from 85 to 476 ms (control 32 ms). Transmission time was lower in OBCs with a larger inner diameter. Shortening the OBCs resulted in a further decrease in transmission time. The type of OBC used has an impact on signal processing. An OBCs with a short transmission time should be chosen, especially if fast pressure changes are to be evaluated such as during phrenic nerve stimulation.