Surface roughness parameters as predictors of anchorage strength in bone: a critical analysis.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 10899340.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The surface roughness of a bone implant was defined parametrically. The values of the parameters defining the surface were varied. Some traditionally used surface roughness parameters were calculated. By means of a theoretical model the bone-implant interfacial shear strength was estimated. No simple correlation between the values of the surface roughness parameters and the estimated interfacial shear strength was found. It was concluded that the value of the traditional surface roughness parameters as predictors of interfacial shear strength is limited. If however a change of the surface topography of an implant is restricted to scale a positive correlation was found between the theoretical interfacial shear strength and some surface roughness parameters. It is suggested that the bone-implant interfacial shear strength in the general case be estimated by means of strength analyses based upon a study of the size, shape and density of the individual elements constituting the rough surface.
Medical subject headings
- Bone and Bones
- Models, Theoretical
- Prostheses and Implants