Unexpected corneal endothelial cell decompensation after intraocular surgery with instruments sterilized by plasma gas.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10919908.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ten cases of unexpected corneal endothelial cell decompensation occurring after routine intraocular surgery using instruments sterilized with a new plasma gas protocol are described. A retrospective observational case series with 1 year of follow-up was conducted. All patients had corneal decompensation and nonreactive pupils after surgery. Six patients required penetrating keratoplasty. Three patients partially recovered pupillary function. Visual acuity at 1 year ranged from 20/20 to hand motion (HM). One patient with an anterior chamber intraocular lens (ACIOL) experienced optic atrophy and HM vision despite resolution of corneal edema. Toxic corneal endothelial cell destruction syndrome was associated with the introduction of plasma gas sterilization protocols.
Medical subject headings