A syringe adapter to facilitate aspiration at amniocentesis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 10928904.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Syringes were developed for injection, not aspiration. We describe an adapter that attaches to a standard syringe and aspirates by advancement rather than withdrawal. The adapter consists of a syringe barrel extension and an attachable finger grip. The adapter attached to a standard syringe allows aspiration by advancement rather than withdrawal, with the same hand motion used as with injection. In a comparison of timed in vitro fluid aspiration by 10 practitioners, mean times for aspiration of 15 mL were 32.4 seconds (range 24-45) with standard technique and 25.3 seconds (range 24-30) with the adapted syringe (P < .003, paired t test). Thirteen of 14 physicians who used the adapted syringe in vivo reported that much less effort was required, and one physician reported that a little less effort was required. Overall, eight practitioners reported that the adapted syringe made amniocentesis much easier and six found no difference between the two devices. This adapter converts a standard syringe to one permitting aspiration by advancement rather than withdrawal. Practitioner feedback was positive with regard to ease of use, degree of effort, and overall satisfaction.
Medical subject headings
- Amniocentesis
- Suction
- Syringes