Wear of high-dose gamma-irradiated polyethylene in total hip replacements.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 10982661.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Between 1971 and 1978, 62 patients underwent total hip replacements (THRs) with high-dose (100 Mrad) gamma-irradiated polyethylene sockets. After a mean follow-up of 17. 3 years (range, 6-23 years), 28 hips in 26 patients were available for radiographic wear measurement. Twenty-three THRs with non-irradiated sockets were used as controls, although all these hips had loosened within 13 years. The rate of linear wear in irradiated sockets was significantly lower throughout the follow-up period (less than 2 years, 0.16 and 0. 38 mm/year; 2-6 years, 0.05 and 0.29 mm/year; more than 6 years, 0. 05 and 0.28 mm/year, respectively; mean, P < 0.0001). This result has demonstrated that high-dose gamma-irradiated, cross-linked polyethylene is a promising articulating material for THRs.