A comparison of anaesthetic tensions in arterial blood and oxygenator exhaust gas during cardiopulmonary bypass.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 10995139.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study evaluates the usefulness of the analysis of gas sampled from the exhaust port of a membrane oxygenator in the estimation of anaesthetic tension in arterial blood. Sixty-seven arterial blood samples were drawn from patients undergoing hypothermic cardiopulmonary bypass with anaesthesia maintained by either isoflurane or desflurane. Anaesthetic tensions in the oxygenator exhaust gas were measured using an infrared analyser and in arterial blood using a two-stage headspace technique with a gas chromatograph. Both measurement systems were calibrated with the same standard gas mixtures. There was no difference in anaesthetic tension measured in arterial blood and gas leaving the oxygenator exhaust (isoflurane: n = 29, range: 0.3-0.8%, 95% limits of agreement: -0.08% to 0.09%; desflurane: n = 38, range: 1.5-5.4%; 95% limits of agreement -0.65% to 0.58%). We conclude that anaesthetic tensions in arterial blood can be accurately monitored by analysis of the gas emerging from the exhaust port of a membrane oxygenator.
Medical subject headings
- Anesthetics, Inhalation
- Cardiopulmonary Bypass
- Monitoring, Intraoperative
- Oxygenators, Membrane