An improved configuration for the reduction of EMG in electrode cuff recordings: a theoretical approach.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11008430.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A theoretical investigation of different electroneurogram recording techniques using electrode cuffs is presented. A new screened tripole arrangement is proposed with a higher inherent signal to interference ratio than the true tripole, which also allows the nulling of the residual electromyogram signal. The reduction in interference is small because the electrode impedance is large compared to the source resistance.
Medical subject headings
- Electrodiagnosis
- Peripheral Nerves