A prospective comparison of three systems of patient immobilization for prostate radiotherapy.

Malone, S; Szanto, J; Perry, G; Gerig, L; Manion, S; Dahrouge, S; Crook, J · Int J Radiat Oncol Biol Phys · 2000

prospective_cohort · Level II

Where this comes from

Abstract

The study compared the setup reliability of 3 patient immobilization systems, a rubber leg cushion, the alpha cradle, and the thermoplastic Hipfix device, in 77 patients with cT1-T3, N0, M0 prostate cancer receiving conformal radiotherapy. Port films were analyzed and compared to simulation films to estimate the setup errors in the three coordinate axes (anterior-posterior, cranial-caudal, medial-lateral). A total vector error was calculated from these shifts. The Hipfix was found significantly superior to the other two devices in reducing mean setup errors in all axes (p < 0.005). The average field-positioning error with the Hipfix ranged from 1.9 mm to 2.6 mm for all axes, whereas the deviation for the other two systems ranged from 2.7 to 3. 4 mm. Errors greater than 10 mm were virtually eliminated with the Hipfix system. There was a reduction in the mean total vector error in the alpha cradle and Hipfix patient cohorts over time, reflecting improved efficacy as a result of experience. There was a significant difference in the performance of each immobilization device. The Hipfix was consistently more reliable in reducing setup errors than the other devices.