Complement activation by PVA as measured by ELIFA (enzyme-linked immunoflow assay) for SC5b-9.

Black, J P; Sefton, M V · Biomaterials · 2000

basic_science · Level V

Where this comes from

Abstract

Polyvinyl alcohol (PVA) coated onto polyethylene (PE) tubes exposed to human serum for 1 hour at 37 degrees C resulted in the production of 1.03 +/- 0.04 microg/cm2 of the soluble form of the terminal membrane attack complex, SC5b-9. This was approximately 20 x that produced by the polyethylene. About one quarter of this total was found associated with the surface of PVA. SC5b-9 concentrations were determined by enzyme-linked immunoflow assay (ELIFA) a variant on ELISA that involved drawing the test sample, the antibodies and the chromogenic reagent through a nitrocellulose membrane filter. ELIFA enabled analysis of protein concentrations in the presence of SDS, so that SDS (0.05%) was used to desorb adsorbed SC5b-9 prior to analysis together with SC5b-9 in the bulk to get a more complete picture of PVA-associated complement activation.