The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 11069336.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fluid delivery from four types of commercially available 50-ml syringes was measured using an electronic balance at an infusion rate of 1 ml.h(-1). Retrograde aspiration volume and zero-drug delivery time were recorded after lowering the syringe pump by 50 cm. Syringe compliance was calculated from the volume of bolus released after occlusion at 100 mmHg. Zero-drug delivery times differed significantly between syringes, ranging from [mean (SD)] 3.26 (0.40) min to 6.38 (0.56) min (F = 55.5, d.f. = 3/20, p < 0.0001). Syringe compliance correlated well with aspiration volume (Pearson r(2) = 0.92, p < 0.001) and zero-drug delivery time (r(2) = 0.90, p < 0.001). Syringe design affected the internal syringe compliance. All syringes were associated with potentially relevant zero-drug delivery times after moderate vertical displacement. To minimise this risk, vertical displacement of syringe pumps delivering highly vasoactive drugs should be avoided.
Medical subject headings
- Drug Delivery Systems
- Syringes