Somatosensory evoked potentials for closed-loop control of anaesthetic depth using propofol in the urethane-anaesthetized rat.

Angel, A; Arnott, R H; Linkens, D A; Ting, C H · Br J Anaesth · 2000

basic_science · Level V

Where this comes from

Abstract

Primary somatosensory cortical mass responses have been shown to exhibit dose-dependent changes in latency when general anaesthetics are administered. Here we describe a system in which the latency of evoked responses was measured automatically in real time in five animals. Latency changes were used to operate a closed-loop control of propofol delivery by intravenous infusion. The system attempted to induce and maintain a 1 ms increase in evoked response latency; this was reversed when infusion was discontinued. Allowing for the rapid and large biological fluctuations in the evoked response, this was achieved successfully. The system maintained a mean increase in latency of 1.27 (SD 0.42) ms. The mean statistical dispersion index of data obtained during the controlled period was 1.23 (0.3); in an ideal controllable system it approximates to 1. Such a system may provide a means for the automatic delivery of anaesthetics.

Medical subject headings