Changes in wall mechanics after endovascular stenting in the rabbit aorta: comparison of three stent designs.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11222230.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to evaluate changes in the wall mechanics of small-diameter arteries after endovascular placement of three different stents. Five self-expandable stents (Wallstent), five balloon-expandable noncovered Palmaz stents, and three balloon-expandable covered stents (Jostent) were placed in the infrarenal aorta of 13 New Zealand white rabbits. Systolic blood pressure changes, blood-flow velocity, systolic diameter, and diameter changes were measured and used to calculate the diameter compliance, the distensibility coefficient, and the pulsatility index. Compliance (10(-3) mm kPa(-1)) was 75.3 +/- 20.1 before stenting and reached 94.7 +/- 42.2 upstream, 38.8 +/- 14.2 at the stent level (p < 0.05), and 70.8 +/- 23.2 downstream from the stent. Distensibility (10(-3) kPa(-1)) was 24.3 +/- 6.3 before stenting and reached 27.8 +/- 10.3 upstream, 10.5 +/- 4.4 at the stent level (p < 0.001), and 21.9 +/- 8.6 downstream from the stent. Compliance and distensibility were significantly lower at the stent level than upstream and downstream (p < 0.05). Aortic diameter increased significantly at the stent level from 3.11 +/- 0.40 mm before to 3.76 +/- 0.42 mm after stenting. No significant difference was found among the three stent designs for all the studied data. Regardless of the three tested stent designs, endovascular stenting produces a significant decrease in arterial wall compliance of the rabbit aorta.
Medical subject headings
- Aorta, Abdominal
- Stents