Radiologic analysis of the tibial intramedullary canal in Chinese varus knees: implications in total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 11222896.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Seventy-two lower limb long radiographs were reviewed with respect to mechanical and anatomic axes. A template of an intramedullary tibial guide rod was employed to determine the accuracy of the intramedullary guide in producing ideal tibial cuts (ie, 90 degrees ) and acceptable tibial cuts (ie, 90 degrees +/- 2 degrees ). The mean difference of the angle formed by the 2 axes was 1.84 degrees +/- 1.42 degrees. In our findings, 22.2% of patients would have unacceptable cuts if an intramedullary device were employed for the tibial cut during a total knee arthroplasty. A radiograph showing the whole tibia is required preoperatively to identify varus tibiae that are not suitable for the intramedullary method.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Tibia
Anatomy
- knee
- tibia