Radiologic analysis of the tibial intramedullary canal in Chinese varus knees: implications in total knee arthroplasty.

Ko, P S; Tio, M K; Ban, C M; Mak, Y K; Ip, F K; Lam, J J · J Arthroplasty · 2001

retrospective_cohort · Level III

Where this comes from

Abstract

Seventy-two lower limb long radiographs were reviewed with respect to mechanical and anatomic axes. A template of an intramedullary tibial guide rod was employed to determine the accuracy of the intramedullary guide in producing ideal tibial cuts (ie, 90 degrees ) and acceptable tibial cuts (ie, 90 degrees +/- 2 degrees ). The mean difference of the angle formed by the 2 axes was 1.84 degrees +/- 1.42 degrees. In our findings, 22.2% of patients would have unacceptable cuts if an intramedullary device were employed for the tibial cut during a total knee arthroplasty. A radiograph showing the whole tibia is required preoperatively to identify varus tibiae that are not suitable for the intramedullary method.

Medical subject headings

Anatomy