An interlaboratory study of a candidate reference method for platelet counting.

Harrison, P; Ault, K A; Chapman, S; Charie, L; Davis, B; Fujimoto, K; Houwen, B; Kunicka, J et al. · Am J Clin Pathol · 2001

Level III

Where this comes from

Abstract

A multinational interlaboratory task force explored the important variables of platelet reference counting and developed a candidate flow cytometric reference method based on the RBC/platelet ratio. A multicenter comparison was performed to determine whether the method met the necessary criteria and was precise enough to be recommended as a new reference method. Each laboratory analyzed serial dilutions of normal specimens, stabilized material, and at least 60 patient specimens with a range of platelet counts from 1 to 400 x 10(3)/microL (1-400 x 10(9)/L). Pooled analysis of the serial dilutions showed that RBC-platelet and RBC-RBC coincidence events became negligible at sufficiently high dilutions (i.e., > 1:1,000). All laboratories demonstrated excellent intra-assay and acceptable interlaboratory precision. Two antibodies (CD61 and CD41) were used for identifying platelets and individually gave acceptable results, but in a minority of samples, staining differences were observed. The optimum method thus uses a double-labeling procedure with a final dilution factor of 1:1,000. The study demonstrated that this method meets the criteria for a reference platelet count.

Medical subject headings