The validity and reliability of the GAITRite system's measurements: A preliminary evaluation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11245768.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the concurrent validity and reliability of the GAITRite computerized gait analysis system with validated paper-and-pencil and video-based methods. Within-groups, repeated-measures design. Research laboratory in a physical therapy education program. One healthy woman, age 27 years. A subject walked across the walkway of the GAITRite system at various walking rates and degrees of step symmetry for 2 of the 3 analyses. Paper placed over the walkway enabled concurrent paper-and-pencil analysis. The subject was concurrently videotaped from the side. For the other analysis, a stride simulator with known step and stride lengths was applied to the walkway to simulate 2 steps and 1 stride. Cadence, walking speed, right and left step and stride lengths, and right and left step times. Excellent paper-and-pencil and GAITRite correlations (intraclass correlation coefficient [ICC] > 95) for spatial measures and excellent video-based and GAITRite correlations (ICC > 93) for temporal measures were found. GAITRite measures of step lengths and times were reliable in both walkway center and left-of-center measurements. Based on this data, GAITRite is a valid and reliable tool for measuring selected spatial and temporal parameters of gait.