Liquid-filled balloon brachytherapy using (68)Ga is effective and safe because of the short 68-minute half-life: results of a feasibility study in the porcine coronary overstretch model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11282912.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
BACKGROUND: Liquid-filled balloons for coronary brachytherapy provide significant advantages over solid sources in dose homogeneity but carry the risk of life-threatening radiointoxication after balloon rupture and laboratory contamination in case of a spill. We hypothesized that the positron emitter (68)Ga, with a half-life of only 68 minutes, was well suited to overcome these safety obstacles while providing full therapeutic efficacy. METHODS AND RESULTS: The feasibility, efficacy, and safety of (68)Ga liquid-filled balloon brachytherapy were investigated in the porcine coronary overstretch model. Four groups of 5 balloon-induced coronary lesions were irradiated with 8, 12, 16, and 24 Gy targeted to the adventitia. Ten unirradiated lesions served as controls. Segments treated with 16 or 24 Gy exhibited marked suppression of neointimal proliferation at 28-day follow-up, with quantitative parameters of intraluminal proliferation reduced to <20%. This beneficial effect was not compromised by untoward edge effects. Uninjured but irradiated vessels did not show histological signs of radiation damage. The (68)Ga whole-body dose due to balloon rupture was estimated to be 5 rem/50 mCi treatment activity and compared favorably with that of (188)Re (78 rem/50 mCi). CONCLUSIONS: (68)Ga positron radiation suppresses neointimal proliferation at doses of 16 and 24 Gy. This biological efficacy, coupled with the attractive safety profile, suggests the selection of (68)Ga as an attractive isotope for liquid-filled balloon brachytherapy.