Xenon expenditure and nitrogen accumulation in closed-circuit anaesthesia.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11284815.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The high price of xenon has prevented its use in routine, clinic anaesthetic practice. Xenon therefore has to be delivered by closed-circuit anaesthesia. The accumulation of nitrogen is a significant problem within the closed circuit and necessitates flushing, which in turn increases gas expenditure and costs. In previous investigations, nitrogen concentrations between 12% and 16% have been reported in closed-circuit anaesthesia. In order to avoid such nitrogen accumulation, we denitrogenised seven pigs using a non-rebreathing system and connected the animals to a system primed with a xenon/oxygen mixture. In comparison, seven pigs were anaesthetised with xenon using a standard low-flow anaesthetic procedure. Anaesthesia time was 2 h. Nitrogen concentrations in the closed system ranged from 0.08 to 7.04% and were not significantly different from those observed during low-flow anaesthesia. Closed-circuit anaesthesia reduced the xenon expenditure 10-fold compared with low-flow anaesthesia.
Medical subject headings
- Anesthesia, Closed-Circuit
- Anesthetics, Inhalation
- Nitrogen
- Xenon