Dose reduction in evacuation proctography.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 11288848.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The goal of this study was to reduce the patient radiation dose from evacuation proctography. Ninety-eight consecutive adult patients referred for proctography to investigate difficult rectal evacuation were studied using a digital imaging system with either a standard digital program for barium examinations, a reduced dose digital program (both with and without additional copper filtration), or Video fluoroscopy. Dose-area products were recorded for each examination and the groups were compared. All four protocols produced technically acceptable examinations. The low-dose program with copper filtration (median dose 382 cGy cm2) and Video fluoroscopy (median dose 705 cGy cm2) were associated with significantly less dose than other groups (p < 0.0001). Patient dose during evacuation proctography can be reduced significantly without compromising the diagnostic quality of the examination. A digital program with added copper filtration conveyed the lowest dose.